Olympus OMNISCAN X3 Phased Array Flaw Detector

The OmniScan X3 flaw detector is a complete phased array toolbox. Powerful tools, like total focusing method (TFM) images and advanced visualization capabilities, enable you to complete your inspection with greater confidence.

Innovative TFM

- Better flaw imaging brings clarity and resolution to smaller defects
- Image early-stage high-temperature hydrogen attack (HTHA) to detect it when it matters most
- Onboard acoustic influence map (AIM) reflectivity simulator helps you visualize TFM sensitivity and adjust accordingly
- Up to four TFM modes facilitates flaw interpretation and sizing

Improved Phased Array

- Twice as fast as the OmniScan MX2 (pulse repetition frequency)
- Single TOFD menu for accelerated workflow
- Improved fast phased array calibration minimizes frustration
- 800% high amplitude range reduces the need to rescan
- Onboard Dual Linear Array™ and Dual Matrix Array™ probe support accelerates setup creation

Get to Work, Quickly

The onboard scan plan, improved fast calibration, and streamlined user interface help eliminate unnecessary steps so that you can complete your inspection setup in minimal time.If you’re an existing OmniScan user, the transition from the MX2 is fast. If you’re new to phased array ultrasonic testing or TFM, the OmniScan X3 flaw detector is easy to learn.

The Reliability You Trust

- IP65 certified rain and dust proof
- Onboard GPS provides the location of data
- Connect wirelessly to the Olympus Scientific Cloud™ to download the latest software
- Scan without stopping thanks to the 25 GB file size

Specifications

Type

Multigroup, multimode ultrasonic flaw detector
Size (W × H × D) 335 mm x 221 mm x 151 mm (13.2 in. x 8.7 in. x 5.9 in.)
Weight 5.7 kg (12.6 lb) (with 1 battery)
Hard Drive Capacity Internal 64 GB SSD, extendable using an external USB drive
Storage Devices SDHC™ and SDXC™ cards or most standard USB storage devices
Max Onboard File Size 25 GB
GPS Yes (unless specified otherwise for some regions)
Alarms 3
Wireless Connection Yes (USB dongle sold separately as an accessory)
PA Connectors 1 connector
UT Connectors 4 (2 channels P/R) for 32:128PR and 16:128PR; the 16:64PR has the same connectors but only one UT channel active (P1/R1)
Certifications ISO 18563-1:2015
EN12668-1:2010
Display
Type TFT LCD with resistive touch screen
Size 269 mm (10.6 in.)
Resolution 1280 × 768 pixels
Number of Colors 16 million
Viewing Angles Horizontal: −85° to 85°
Vertical: −85° to 85°
I/O Ports
USB 2.0 2 ports (one hidden behind the battery)
USB 3.0 1 port
Video Output Video out (HDMI)
Memory Card SDHC port
Communication Ethernet
I/O Lines
Encoder 2-axis encoder line (quadrature or clock/direction), 3rd encoder ready
Digital Input 6 digital inputs, TTL
Digital Output 5 digital inputs, TTL
Acquisition On/Off Switch Through the configuration of a digital input
Power Output Line 5 V nominal, 1 A (short-circuit protected), and 12 V output at 1 A
External DC Supply
DC-IN Voltage 15 VDC to 18 VDC (min. 50 W)
Connector Circular, 2.5 mm pin diameter, center-positive
Battery
Type Lithium-ion battery
Capacity 93 Wh
Number of Batteries 2
Life 5 hours using 2 batteries (hot-swap capable)
PA/UT Configuration
Bit Depth 16 bits
Maximum PRF 12 kHz
Frequency
Effective Digitizing Frequency Up to 100 MHz
Display
Refresh Rate A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Envelope (echo-dynamic mode) Yes: Volume-corrected S-scan (30 Hz)
A-Scan Height Up to 800%
Synchronization
On Internal Clock 1 Hz to 10 kHz
External Pace Yes
On Encoder On 2 axes: from 1 to 65,536 steps

Data Specifications

Processing
Maximum Number of A-Scan Data Points Up to 16,384
Real-Time Averaging PA: 2, 4, 8, 16
UT: 2, 4, 8, 16, 32, 64
Rectification RF, full wave, half wave+, half wave-
Filtering PA channel: 3 low-pass, 6 band-pass, and 4 high-pass filters
UT channel: 8 low-pass, 6 band-pass, and 4 high-pass filters (3 low-pass filters when configured in TOFD)
Video Filtering Smoothing (adjusted to the probe frequency range)
Programmable TCG
Number of Points 32: One TCG (time-corrected gain) curve per focal law
Range PA (standard): 40 dB per step of 0.1 dB
PA (extended): 65 dB per step of 0.1 dB
UT: 100 dB per step of 0.1 dB
Maximum Slope PA (standard): 40 dB/10 ns
PA (extended): 0.1 dB/10 ns
UT: 40 dB/10 ns

Acoustic Specifications

PulserPA ChannelUT Channels
Voltage 40 V, 80 V, and 115 V 85 V, 155 V, and 295 V
Pulse Width Adjustable from 30 ns to 500 ns; resolution of 2.5 ns Adjustable from 30 ns to 1,000 ns; resolution of 2.5 ns
Fall Time < 10 ns < 10 ns
Pulse Shape Negative square pulse Negative square pulse
Output Impedance 28 Ω in pulse-echo
24 Ω in pitch-catch
< 30 Ω
ReceiverPA ChannelUT Channels
Gain Range 0 dB to 80 dB maximum input signal; 550 mVp-p (full-screen height) 0 dB to 120 dB maximum input signal; 34.5 Vp-p (full-screen height)
Input Impedance 57 Ω ± 10% at 9 MHz in pulse-echo
100 Ω ± 10% at 9 MHz in pitch-catch
50 Ω in pulse-echo mode
50 Ω in pulse-receive mode
System Bandwith 0.5 MHz to 18 MHz 0.25 MHz to 28 MHz
Beam FormationPA ChannelUT Channels
Scan Type Single, linear, sectorial, compound, and TFM -
Maximum Aperture OMNIX3-PATFM1664PR = 16 elements
OMNIX3-PATFM16128PR = 16 elements
OMNIX3-PATFM32128PR = 32 elements
-
Number of Receiving Elements OMNIX3-PATFM1664PR = 64 receiving elements
OMNIX3-PATFM16128PR = 128 receiving elements
OMNIX3-PATFM32128PR = 128 receiving elements
-
Number of Focal Laws Up to 1024 -
Delay Range Transmission 0 µs to 10 µs in 2.5 ns increments -
Delay Range Reception 0 µs to 6.4 µs in 2.5 ns increments -

TFM/FMC

Supported Modes Pulse-echo: L-L, T-T, and TT-TT
Self-tandem: LL-L, TT-T, TT-L, TL-T, LT-T, and TL-L
Number of Groups Up to 4 simultaneous TFM groups
Maximum Aperture OMNIX3-PATFM1664PR = 32 elements extended aperture
OMNIX3-PATFM166128PR = 32 elements extended aperture
OMNIX3-PATFM32128PR = 64 elements extended aperture
Image Resolution Up to 1024 × 1024 (1 M points) for each TFM group
Live TFM Envelope Yes

Operating Environment

Ingress Protection Rating IP65 certified (completely protected against dust and water jets from all directions (6.3 mm nozzle))
Shockproof Rating Drop tested according to MIL-STD-810G
Intended Use Indoor and outdoor use
Altitude Up to 2,000 m (6,561 ft)
Operating Temperature 0 °C to 45 °C (32 °F to 113 °F)
Storage Temperature

−20 °C to 60 °C (−4 °F to 140 °F) (with battery inside)
−20 °C to 70 °C (−4 °F to 158 °F) (with no battery inside)


Package Includes:

- Omniscan X3 Unit
- Battery Pack
- AC Adapter/Charger
- Power Cord
- User Manual On Cd-ROM

- Carrying Case (G)

Warranty Information

2 Year Warranty

Find Similar Products by Tag


Find Similar Products by Category


Write your own product review

Product Reviews

This product hasn't received any reviews yet. Be the first to review this product!


Add to Wish List

Click the button below to add the Olympus OMNISCAN X3 Phased Array Flaw Detector to your wish list.

You Recently Viewed...